Getting Started
Environmental Requirements
18
Environmental Requirements
Keysight Technologies has designed this product for use in
Installation Category II, Pollution Degree 2, per IEC 61010-1.
Keysight has designed the spectrum analyzer for use under the
following conditions:
• Indoor use
• Altitude < 3,000 meters
• Operating temperature range: +5 to +45
o
C;
Storage temperature range: –20 to +70
o
C
• Relative humidity range 15% to 95 % at 40
o
C
Ventilation
Ventilation holes are located on the rear panel and one side of the
spectrum analyzer cover. Do not allow these holes to be
obstructed, as they allow air flow through the spectrum analyzer.
When installing the spectrum analyzer in a cabinet, do not restrict
the convection of the analyzer. The ambient temperature outside
the cabinet must be less than the maximum operating temperature
of the spectrum analyzer by 4
o
C for every 100 watts dissipated
within the cabinet.
Cleaning Tips
To prevent electrical shock, disconnect the spectrum analyzer from
line power before cleaning. Use a dry cloth or one slightly
dampened with water to clean the external case parts. Do not
attempt to clean internally.
NOTE
This equipment is Class A suitable for professional use and is for use in elec-
tromagnetic environments outside of the home.
Содержание N9321C
Страница 36: ...Getting Started Helpful Tips 28 ...
Страница 110: ...Functions and Measurements Files Operation 102 ...
Страница 180: ...Key Reference Trace Det 172 ...
Страница 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 216: ...Menu Maps Trace Det Menu 208 ...
Страница 219: ......