Functions and Measurements
Analysis the Modulated Signals
92
3 Press
Frequency > Carrier Freq > 100 > MHz
to set the carrier
frequency.
4 Press
Meas > Symbol Rate > 10 > ksps
to set the symbol rate.
5 Press
Filter Setup > Ref Filter > Nyquist
to set the filter type.
6 Set the attenuation state to Auto (default):
Press
Amplitude > Attenuation Auto
7 Setting the FSK deviation limit and turn on the limit:
Press
Shift > Limit > FreqDev UP > 85 > kHz > FreqDev Low > 80 >
kHz > Limits on
The figure below shows the FSK symbol and the measurement
result.
Figure 3-32 FSK Symbol View
Содержание N9321C
Страница 36: ...Getting Started Helpful Tips 28 ...
Страница 110: ...Functions and Measurements Files Operation 102 ...
Страница 180: ...Key Reference Trace Det 172 ...
Страница 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 216: ...Menu Maps Trace Det Menu 208 ...
Страница 219: ......