Functions and Measurements
Stimulus Response Transmission
60
5
To increase measurement sensitivity and smooth the noise:
Press
BW > 30 > kHz
Press
BW >
Video BW > 300 > kHz
A
decrease
in
displayed
amplitude
is
caused
by
tracking error
.
6
Connect the cable from the tracking generator output to the
analyzer input. Store the frequency response in trace 3 and
normalize:
Press
Meas > Normalize > Store Ref (1
→
4) > Normalize (On)
7
Reconnect the DUT to the analyzer and change the normalized
reference position:
Press
Meas > Normalize > Norm Ref Posn > 8
8
Measure the rejection of the low pass filter:
Press
Marker > 370 > MHz,
Press
Mode > Delta > 160> MHz
NOTE
To reduce ripples caused by source return loss, use 6 dB or greater
output attenuation. Tracking generator output attenuation is normally a
function of the source power selected. However, the output attenuation
may be controlled in the Tracking Generator menu.