The procedure below describes using the built-in tracking
generator of the analyzer to measure the rejection of a low pass
filter. A type of transmission measurement.
1
To measure the rejection of a low pass filter, connect the
equipment as shown in
. A 370 MHz low pass filter as
the DUT.
Figure 3-18Transmission Measurement Test Setup
2
Press
Preset
. (With
Preset Type
of
DFT
)
3
Set the start and stop frequencies and resolution bandwidth:
Press
Frequency > Start Freq > 100 > MHz
Press
Frequency> Stop Freq > 1 > GHz
Press
Bw > 3 > MHz
4
Turn on the tracking generator and if necessary, set the output
power to -10 dBm:
Press
Mode > Tracking Generator > Amplitude (On) > -10 > dBm
Remote
Standby
On
TG SOURCE
CAL OUT
50M Hz
10dBm
7
·
4
1
0
2
5
8
9
6
3
Back
Enter
Marker
Peak
Search
M arker
Auto
Tune
Det/
Display
File/
BW/
Avg
View/
Trace
M eas
MODE
Sw eep/
Trig
Local
Save
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
PROBE POWER
RF IN 50
50V DC M AX
30dBm 1W MAX
CAT
Ⅱ
DUT
CAUTION
Excessive signal input may damage the DUT. Do not exceed the
maximum power that the device under test can tolerate.
Содержание N9321C
Страница 36: ...Getting Started Helpful Tips 28 ...
Страница 110: ...Functions and Measurements Files Operation 102 ...
Страница 180: ...Key Reference Trace Det 172 ...
Страница 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 216: ...Menu Maps Trace Det Menu 208 ...
Страница 219: ......