Functions and Measurements
Analysis the Modulated Signals
88
6 Set the AM depth limit and turn on the limit function:
Press
Shift > Limit > AM Depth UP > 51 > % > AM Depth Low > 49
> % > Limits on
The red value in the result indicates the demodulated AM depth
exceeds the limits set as above.
Figure 3-29 AM demodulation analysis
FM Modulation Analysis
Please refer to the procedures below for the FM modulation
analysis:
1 Connect the signal generator to RF IN connector and turn on
the signal generator’s FM and RF output.
2 Turn on the FM demodulation analysis function.
Press
MODE > Modulation Analysis > AM
.
3 Press
Frequency > Carrier Freq > 100 > MHz
to set the carrier
frequency.
4 Set the attenuation state to Auto (default):
Press
Amplitude > Attenuation Auto
5 Press
Shift > Disp > Y Scale
> Scale Coupling On
to set the Y scale
auto coupling.