Key Reference
BW
ratio by using the arrow keys, knob, or numeric keypad. When
VBW/RBW (Auto)
is selected, the ratio is determined automatically
by the rules in
Figure 5VBW/RBW Ratio Auto Rules
Key access:
BW > VBW/RBW
Average Type
Chooses one of the following averaging scales: log power (video),
power (RMS), or voltage averaging. When trace average is on, the
average type is shown on the left side of the display. Selecting an
average type manually causes the analyzer to use that type without
regard to other analyzer settings.
Key access:
BW > Average Type
Is Detector peak
or Negative peak?
Is Trace
Average On?
Is Detector
Sample?
Is Avg Type
Power?
Is Trace
Average On?
Is Avg Type
Power?
Ratio=10
Ratio=1
Ratio=10
Ratio=1
Ratio=10
Ratio=1
Ratio=1
Yes
Yes
No
No
No
No
Yes
Yes
Yes
No
No
Yes