Functions and Measurements
Perform an EMC Measurement
96
9
Press
[Shift] > [Limit/Sweep] > {Limit 1} > {Recall Limit}
to load a
CISPR limit line from memory.
10
Toggle
{Limit On/Off}
to On to set the positive peak detector..
11
Press
[Peak Search]
to place a marker on the signal which is out of
limit.
12
Press
[Marker] > {Marker To} > {To Center}
13
Press
[Span] > {Zero Span}
14
Press
[Trace/Det] > {Detector} > {Quasi peak}
to set the quasi peak
detector.
15
Press
[Marker] > {Marker To} > {To Center}
NOTE
You can edit a limit line file by pressing [Shift] > {Limit} > {Limit 1} > {Edit
limit}, or in HSA and BSA PC software by clicking {File} > {New} > {Limit}