12
SYSTem Subsystem
466
Keysight N8262A Programming Guide
–226
Lists not same length
This occurs when SENSe:CORRection:CSET[1]|CSET2:STATe is set to ON and the frequency and
calibration/offset lists do not correspond in length.
–230
Data corrupt or stale;Please calibrate Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has not been
calibrated, then any command which would normally return a measurement result (for example FETC?,
READ?, or MEAS?) will generate this error message.
–231
Data questionable;CAL ERROR
Power meter calibration failed. The most likely cause is attempting to calibrate without applying a 1 mW
power to the power sensor.
–231
Data questionable;CAL ERROR ChA
Power meter calibration failed on Channel A. The most likely cause is attempting to calibrate without
applying a 1 mW power to the power sensor.
–231
Data questionable;CAL ERROR ChB
Power meter calibration failed on Channel B. The most likely cause is attempting to calibrate without
applying a 1 mW power to the power sensor.
–231
Data questionable;Input Overload
The power input to Channel A exceeds the power sensor’s maximum range.
–231
Data questionable;Input Overload ChA
The power input to Channel A exceeds the power sensor’s maximum range.
–231
Data questionable;Input Overload ChB
The power input to Channel B exceeds the power sensor’s maximum range.
–231
Data questionable;Lower window log error
This indicates that a difference measurement in the lower window has given a negative result when the
units of measurement were logarithmic.
–231
Data questionable;Upper window log error
This indicates that a difference measurement in the upper window has given a negative result when the
units of measurement were logarithmic.
–231
Data questionable;ZERO ERROR
Power meter zeroing failed. The most likely cause is attempting to zero when some power signal is being
applied to the power sensor.
–231
Data questionable;ZERO ERROR ChA
Power meter zeroing failed on Channel A. The most likely cause is attempting to zero when some power
signal is being applied to the power sensor.
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Страница 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 48: ...48 Keysight N8262A Programming Guide ...
Страница 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...1 Power Meter Remote Operation 122 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 184: ...2 MEASurement Commands 184 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...3 CALCulate Subsystem 222 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 274: ...5 DISPlay Subsystem 274 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 326: ...8 OUTPut Subsystem 326 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 340: ...9 PSTatistic Subsystem 340 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 418: ...10 SENSe Subsystem 418 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 444: ...11 STATus Subsystem 444 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 536: ...12 SYSTem Subsystem 536 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 564: ...13 TRACe Subsystem 564 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Страница 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Страница 650: ...16 SERVice Subsystem 650 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...
Страница 672: ...17 IEEE 488 2 Command Reference 672 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...