Keysight N8262A Programming Guide
Getting the Best Speed Performance
This section discusses the factors that influence the speed of operation (number
of readings/sec) of a P-Series modular power meter.
The following factors are those which have the greatest effect upon measurement
speed (in no particular order):
– The selected measurement rate, i.e.
NORMal
,
DOUBle
,
FAST
.
– The sensor being used.
– The trigger mode (for example, free run, trigger with delay etc.).
– The output format:
ASCii
or
REAL
.
– The units used for the measurement.
– The command used to take a measurement.
In addition, in
FAST
mode there are other influences which are described in
The following paragraphs give a brief description of the above factors and how
they are controlled from SCPI.
Measurement Rate
There are three possible speed settings
NORMal
,
DOUBle
and
FAST
. These are set
using the
SENSe:MRATe
command and can be applied to each channel
independently.
In
NORMal
and
DOUBle
modes, full instrument functionality is available and these
settings can be used with all sensors.
FAST
mode is only available for the P-Series
and E-Series sensors. Also, in FAST mode averaging, limits and ratio/difference
math functions are disabled.
Refer to “Specifications” in the
P-Series Modular Power Meter User’s Guide
to see
the influence of these speed settings on the accuracy and noise performance of
the power meter.
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Страница 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 48: ...48 Keysight N8262A Programming Guide ...
Страница 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...1 Power Meter Remote Operation 122 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 184: ...2 MEASurement Commands 184 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...3 CALCulate Subsystem 222 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 274: ...5 DISPlay Subsystem 274 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 326: ...8 OUTPut Subsystem 326 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 340: ...9 PSTatistic Subsystem 340 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 418: ...10 SENSe Subsystem 418 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 444: ...11 STATus Subsystem 444 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 536: ...12 SYSTem Subsystem 536 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 564: ...13 TRACe Subsystem 564 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Страница 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Страница 650: ...16 SERVice Subsystem 650 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...
Страница 672: ...17 IEEE 488 2 Command Reference 672 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...