4
CALibration Subsystem
230
Keysight N8262A Programming Guide
CALibration[1]|2:AUTO [ONCE|ON|OFF|0|1]
This command calibrates the specified channel when enabled. The command
assumes that an 8480 or E-Series power sensor is connected to a 1 mW reference
signal.
1|ON
can only be used with a P-Series sensor. When
1|ON
is enabled the
calibration is updated if the meter’s or sensor’s temperature changes by
±
5
o
C or
the time since last calibration is greater then 1000 minutes.
The
0|OFF
parameter is only required for the query response and is ignored in the
command.
The E-Series power sensors have their sensor calibration tables stored in EEPROM
which means that the reference calibration factor is automatically downloaded by
the power meter.
For 8480 Series power sensors the reference calibration factor used during this
calibration can be obtained from an active sensor calibration table or the value
entered using
CALibration:RCFactor
. The actual value used is the one which
was most recently set. For example, a value entered using
CALibration:RCFactor
is overridden if a sensor calibration table is subsequently selected and enabled
and
CALibration:RCFactor
overrides any reference calibration factor previously
set from a sensor calibration table. To determine the current reference calibration
factor, use
CALibration:RCFactor?
.
NOTE
If the power meter is using an 8480 or E-Series power sensor it should be zeroed
before calibration using the
CALibration:ZERO:AUTO ONCE
command.
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Страница 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 48: ...48 Keysight N8262A Programming Guide ...
Страница 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...1 Power Meter Remote Operation 122 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 184: ...2 MEASurement Commands 184 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...3 CALCulate Subsystem 222 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 274: ...5 DISPlay Subsystem 274 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 326: ...8 OUTPut Subsystem 326 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 340: ...9 PSTatistic Subsystem 340 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 418: ...10 SENSe Subsystem 418 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 444: ...11 STATus Subsystem 444 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 536: ...12 SYSTem Subsystem 536 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 564: ...13 TRACe Subsystem 564 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Страница 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Страница 650: ...16 SERVice Subsystem 650 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...
Страница 672: ...17 IEEE 488 2 Command Reference 672 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...