SYSTem Subsystem
12
Keysight N8262A Programming Guide
473
SYSTem:PRESet <character_data>
This command presets the power meter to values appropriate for measuring the
communications format specified by <character_data>. The power meter is preset
to default values if no value or the value DEFault is supplied.
For further information on preset configurations, refer to
through to
Command results differ according to the sensor(s) connected to the power meter:
– If connected to a power meter, a P-Series sensor are connected to
one channel, and another model sensor is connected to the other channel, the
channel connected to the P-Series sensor is set up according to the
<character_data> value and the other channel is set to DEFault values.
– If two P-Series sensors are connected to power meter, both channels are set to
the same values except for bandwidth which is set to an appropriate value for
each sensor.
Primary and Secondary Channels
P-Series modular power meter channels are defined as either primary or
secondary. The primary channel is always the trigger master and primary channel
measurements occupy a greater share of the display space than secondary
channel measurements.
– If a power meter has a P-Series sensor connected, the P-Series sensor channel
is the primary channel. In such cases the primary channel could be either
Channel A or Channel B. The other model’s channel is the secondary channel.
– If a power meter has two, P-Series sensors, connected to it, the primary
channel is always Channel A and the secondary channel is Channel B.
NOTE
DEFault settings apply to both
*RST
and to
SYSTem:PREset DEFault
unless
stated otherwise.
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Страница 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 48: ...48 Keysight N8262A Programming Guide ...
Страница 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...1 Power Meter Remote Operation 122 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 184: ...2 MEASurement Commands 184 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...3 CALCulate Subsystem 222 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 274: ...5 DISPlay Subsystem 274 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 326: ...8 OUTPut Subsystem 326 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 340: ...9 PSTatistic Subsystem 340 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 418: ...10 SENSe Subsystem 418 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 444: ...11 STATus Subsystem 444 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 536: ...12 SYSTem Subsystem 536 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 564: ...13 TRACe Subsystem 564 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Страница 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Страница 650: ...16 SERVice Subsystem 650 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...
Страница 672: ...17 IEEE 488 2 Command Reference 672 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...