10
SENSe Subsystem
370
Keysight N8262A Programming Guide
Example
Reset Condition
On reset the selected table is not affected.
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect]?
The name of the selected table is returned as a quoted string. If no table is
selected an empty string is returned.
Query Example
Error Messages
– If
<string>
is not valid, error –224, “Illegal parameter value” occurs.
– If a table called
<string>
does not exist, error –256, “File name not found”
occurs.
– When a sensor calibration table is selected, the power meter verifies that the
number of calibration points defined is one more than the number of frequency
points defined. When a frequency dependent offset table is selected, the
power meter verifies that the number of offset points defined is equal to the
number of frequency points defined. If this is not the case, error –226, “Lists
not the same length” occurs.
– If the
CSET1
command is used when a P-Series or an E-Series Power Sensor is
connected the error –241, “Hardware missing” occurs.
CORR:CSET1 ‘PW1’
This command enters the name of the sensor
calibration table which is to be used on
Channel A.
CORR:CSET1?
This command queries the sensor calibration
table currently used for Channel A.
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Страница 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 48: ...48 Keysight N8262A Programming Guide ...
Страница 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...1 Power Meter Remote Operation 122 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 184: ...2 MEASurement Commands 184 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...3 CALCulate Subsystem 222 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 274: ...5 DISPlay Subsystem 274 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 326: ...8 OUTPut Subsystem 326 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 340: ...9 PSTatistic Subsystem 340 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 418: ...10 SENSe Subsystem 418 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 444: ...11 STATus Subsystem 444 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 536: ...12 SYSTem Subsystem 536 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 564: ...13 TRACe Subsystem 564 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Страница 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Страница 650: ...16 SERVice Subsystem 650 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...
Страница 672: ...17 IEEE 488 2 Command Reference 672 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...