UNIT Subsystem
15
Keysight N8262A Programming Guide
607
UNIT[1]|2|3|4:POWer <amplitude_unit>
This command sets the power measurement units for a specified window/
measurement. The power suffix set by
UNIT:POWer
is used for any command
which accepts a numeric value in more than one unit
–
UNIT1:POWer
sets the power measurement units for the upper window/upper
measurement.
–
UNIT2:POWer
sets the power measurement units for the lower window/upper
measurement.
–
UNIT3:POWer
sets the power measurement units for the upper window/lower
measurement.
–
UNIT4:POWer
sets the power measurement units for the lower window/lower
measurement.
For ratio and relative power measurements:
– If
UNIT:POWer
is W, the measurement units are percentage.
– If
UNIT:POWer
is DBM, the measurement units are dB relative.
Syntax
amplitude_unit
:POW
Space
?
UNIT
1
2
3
4
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Страница 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 48: ...48 Keysight N8262A Programming Guide ...
Страница 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...1 Power Meter Remote Operation 122 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 184: ...2 MEASurement Commands 184 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...3 CALCulate Subsystem 222 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 274: ...5 DISPlay Subsystem 274 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 326: ...8 OUTPut Subsystem 326 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 340: ...9 PSTatistic Subsystem 340 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 418: ...10 SENSe Subsystem 418 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 444: ...11 STATus Subsystem 444 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 536: ...12 SYSTem Subsystem 536 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 564: ...13 TRACe Subsystem 564 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Страница 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Страница 650: ...16 SERVice Subsystem 650 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...
Страница 672: ...17 IEEE 488 2 Command Reference 672 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...