104
90000 A-Series Service Guide
5
Replacing Assemblies
2
Perform a service calibration after the warm-up period:
a
Connect the RS-232 cable to the Digital Multimeter RS-232 port and to the
RS-232 port on the oscilloscope.
b
Set the Digital Multimeter as follows:
c
Click
Utilities > Calibration
.
d
Uncheck the Cal Memory Protect box to allow calibration.
e
Click
Start
to start the calibration procedure.
f
Follow the on-screen instructions as calibration proceeds.
Interface: RS-232
Baud Rate: 9600 Baud
Parity: None: 8 bits
Language: SCPI