62
90000 A-Series Service Guide
4
Troubleshooting
Over-Voltage Fault
If the oscilloscope loses power almost immediately as described earlier, it is
probably an over-voltage fault.
Three groups of LEDs indicate over-voltage faults. One group is located near the
center of the top edge of the backplane assembly just above the blue and red
SATA cables.
If any of these LEDs are lit, the backplane assembly must be replaced.
The other two groups of LEDs that indicate over-voltage faults are located along
the edges of the two acquisition assemblies and are visible through the bottom fan
opening located on the left side of the oscilloscope, nearest the rear panel.
Содержание Infiniium 90000A Series
Страница 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Страница 6: ...6 90000 A Series Oscilloscopes Service Guide...
Страница 12: ...12 90000 A Series Service Guide 1 General Information...
Страница 18: ...18 90000 A Series Service Guide 2 Calibration...
Страница 110: ...110 90000 A Series Service Guide 5 Replacing Assemblies P owerharness between backplaneand m otherboard O n O ff ca ble...
Страница 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Страница 119: ...Replaceable Parts 6 90000 A Series Service Guide 119 Exploded Views Front Frame and Front Panel...
Страница 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Страница 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Страница 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Страница 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Страница 135: ......