Testing Performance
3
90000 A-Series Service Guide
29
Procedure
1
Make the connections to oscilloscope channel 1 as shown below.
NOTE
•
Where the BNC Tee adapter is used, it is important to connect it directly to the oscilloscope
channel input using the BNC (f) to SMA (m) adapter and the connector saver to minimize
ground potential differences and to ensure that the DMM measures the input voltage to the
oscilloscope channel as accurately as possible. Differences in ground potential can be a
significant source of measurement error, particularly at high scope sensitivities.
•
It also helps to reduce ground potential differences if the oscilloscope and DMM are
connected to the same AC supply circuit.
•
256 averages are used in the oscilloscope measurements of this section to reduce
measurement noise and to reduce the measurement error due to resolution.
Содержание Infiniium 90000A Series
Страница 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Страница 6: ...6 90000 A Series Oscilloscopes Service Guide...
Страница 12: ...12 90000 A Series Service Guide 1 General Information...
Страница 18: ...18 90000 A Series Service Guide 2 Calibration...
Страница 110: ...110 90000 A Series Service Guide 5 Replacing Assemblies P owerharness between backplaneand m otherboard O n O ff ca ble...
Страница 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Страница 119: ...Replaceable Parts 6 90000 A Series Service Guide 119 Exploded Views Front Frame and Front Panel...
Страница 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Страница 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Страница 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Страница 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Страница 135: ......