Testing Performance
3
90000 A-Series Service Guide
41
12
Convert the average frequency value to time scale error by subtracting 20 Hz
and dividing by 10 Hz/ppm. Record the result in the Measured Time Scale Error
(ppm) column of the Time Scale Accuracy table.
13
Record the time since calibration (in years) in the table. The calibration date
can be found in the Calibration window (
Utilities > Calibration...
).
14
Calculate the test limits using the following formula and record them in the
table.
Test Limits = ±(0.4 + 0.5 x Years Since Calibration)
15
Record the results in the Performance Test Record.
Содержание Infiniium 90000A Series
Страница 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Страница 6: ...6 90000 A Series Oscilloscopes Service Guide...
Страница 12: ...12 90000 A Series Service Guide 1 General Information...
Страница 18: ...18 90000 A Series Service Guide 2 Calibration...
Страница 110: ...110 90000 A Series Service Guide 5 Replacing Assemblies P owerharness between backplaneand m otherboard O n O ff ca ble...
Страница 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Страница 119: ...Replaceable Parts 6 90000 A Series Service Guide 119 Exploded Views Front Frame and Front Panel...
Страница 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Страница 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Страница 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Страница 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Страница 135: ......