42
90000 A-Series Service Guide
3
Testing Performance
Performance Test Record
Keysight Technologies
Keysight 90000 V-Series Oscilloscopes
Model Number _____________________
Tested by________________________
Serial Number ___________________________
Work Order No.___________________
Recommended Test Interval—1 Year/2000 hours
Date_____________________________
Recommended next test date ___________________
Ambient temperature ______________
Offset Accuracy—Zero Error Test
Vertical Scale
Test Limits
Channel 1
Channel 2
Channel 3
Channel 4
5 mV/div
-1.4 mV to +1.4 mV
10 mV/div
-1.8 mV to +1.8 mV
20 mV/div
-2.6 mV to +2.6 mV
50 mV/div
-5.0 mV to +5.0 mV
100 mV/div
-9.0 mV to +9.0 mV
200 mV/div
-17.0 mV to +17.0 mV
500 mV/div
-41.0 mV to +41.0 mV
1 V/div
-81.0 mV to +81.0 mV
Содержание Infiniium 90000A Series
Страница 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Страница 6: ...6 90000 A Series Oscilloscopes Service Guide...
Страница 12: ...12 90000 A Series Service Guide 1 General Information...
Страница 18: ...18 90000 A Series Service Guide 2 Calibration...
Страница 110: ...110 90000 A Series Service Guide 5 Replacing Assemblies P owerharness between backplaneand m otherboard O n O ff ca ble...
Страница 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Страница 119: ...Replaceable Parts 6 90000 A Series Service Guide 119 Exploded Views Front Frame and Front Panel...
Страница 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Страница 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Страница 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Страница 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Страница 135: ......