Replacing Assemblies
5
90000 A-Series Service Guide
89
6
Disengage the tabs located along the edges of the control panel faceplate.
Three tabs are located along the right side (first photo below) and two more
tabs are on the top and bottom of the left edge (following two photos). You
need to look inside the instrument to see the left side tabs. You can see the top
left side tab by looking through the top cavity, and the bottom left tab by
turning the oscilloscope on its side and looking through the bottom cavity.