DDR2(+LP) Compliance Testing Methods of Implementation
129
Single-Ended Signals AC parameter tests for Clocks
8
Measurement Algorithm
1 Pre-condition the oscilloscope.
2 Trigger on a rising edge of the clock signal under test.
3 Find all valid Clock positive pulses in the entire waveform. A valid Clock positive pulse starts at
the V
REF
crossing on a valid Clock rising edge and ends at the V
REF
crossing on the following
valid Clock falling edge.
4 For the first valid Clock positive pulse, zoom in on the pulse so that it appears on the
oscilloscope’s display and perform T
MAX
. Then perform V
TIME
at the found T
MAX
to get the
maximum voltage of the pulse. Take the V
TIME
measurement result as the V
SEH(AC)
value.
5 Continue the previous step for the rest of the valid Clock positive pulses that were found in the
waveform.
6 Determine the worst result from the set of V
SEH(AC)
measured.
Содержание D9020DDRC
Страница 1: ...Keysight D9020DDRC DDR2 LP Compliance Test Application Methods of Implementation ...
Страница 10: ...10 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 40: ...1 Installing the DDR2 Compliance Test Application 40 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 46: ...2 Preparing to Take Measurements 30 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 70: ...3 Measurement Clock Tests 54 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 134: ...6 Single Ended Signals VIH VIL Data Mask Tests 118 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 158: ...9 Single Ended Signals Overshoot Undershoot Tests 142 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 186: ...10 Differential Signals AC Input Parameters Tests 170 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 224: ...14 Clock Timing CT Tests 208 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 270: ...15 Data Strobe Timing DST Tests 254 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 342: ...17 Command and Address Timing CAT Tests 326 DDR2 LP Compliance Testing Methods of Implementation ...
Страница 366: ...19 Calibrating the Infiniium Oscilloscope and Probe 350 DDR2 LP Compliance Testing Methods of Implementation ...