Getting Started
Applying DC Output
This example sets the Source value to +500 mV.
Step 3.
Setting the Limit value (compliance current)
1. Press the
Limit
assist key on a 1-ch model, or the
Ch1 Limit
assist key on a 2-ch
model. Or, use the rotary knob or arrow keys to move the field pointer onto the
Limit value and press the knob. The field pointer appears on the Limit value, and
the status changes to EDIT (green).
2. Use the rotary knob, arrow keys, or numeric/alpha keys
to set the value.
3. Press one of the unit assist keys to set the unit and fix the value.
The status changes to MOVE (blue).
This example sets the Limit value to 1 mA.
Step 4.
Enabling the channel
Press the Ch 1
switch to enable channel 1. This turns the switch green.
Channel 1 starts applying the voltage specified by the Source value.
Step 5.
Disabling the channel
Press the Ch 1
switch to disable channel 1. This turns off the switch light.