5-18
Keysight B2900 User’s Guide, Edition 5
Front Panel Operations
To Set Source Output
To Set Output Filter
For details on this function, see
.
1. Press the Config > Source > Filter function keys.
The Output Filter dialog box opens.
2. Set each parameter.
3. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
To Set Connection Type
For details on this function, see
“2-Wire Connections or 4-Wire Connections” on
1. Press the Config > Source > Connection function keys.
The Output Connection dialog box opens.
2. In the Ch field, set the channel to which the setting change is applied.
3. Set 2-WIRE or 4-WIRE in the Sensing Type field.
4. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
To Set Low Terminal State
For details on this function, see
.
1. Press the Config > Source > Connection function keys.
The Output Connection dialog box opens.
2. In the Ch field, set the channel to which the setting change is applied.
3. Set GROUNDED or FLOATING in the Low Terminal State field.
4. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
Содержание B2900 Series
Страница 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Страница 3: ......
Страница 25: ...1 Getting Started...
Страница 41: ...2 Introduction...
Страница 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Страница 65: ...3 Installation...
Страница 101: ...4 Front Panel Reference...
Страница 155: ...5 Front Panel Operations...
Страница 189: ...6 Function Details...
Страница 233: ......