Keysight B2900 User’s Guide, Edition 5
6-35
Function Details
Initial Settings
Initial Settings
The B2900 is initialized by turning the B2900 on, the *RST command, or the device
clear. Initial settings of the B2900 are shown in this section.
Table 6-6
System Initial Settings
Setup item
Power on
Reset
GPIO function
DIGITAL IN
GPIO function (D14)
HIGH VOLTAGE LAMP
GPIO polarity
NEGATIVE
GPIO output trigger type
EDGE
GPIO output trigger timing
BOTH
GPIO output trigger pulse width
100
s
Display digits
7
Zoomed display
OFF
Not changed
Display image format
JPG
User message enable
Disabled
User message
“”
Byte swap (measurement data)
OFF
Data format (measurement data)
ASCII
Data format (GPIO data)
ASCII
Data format (status register)
ASCII
Data elements (measurement data)
All (V/I/R/S/T/Stat)
Data elements (calculate)
CALC
Automatic timestamp reset
ON
Mass storage directory
(Root)