Keysight B2900 User’s Guide, Edition 5
2-9
Introduction
Source/Measure Unit
Source/Measure Unit
The source/measure unit (SMU) can apply voltage or current, and can measure
current or voltage.
shows a simplified SMU circuit diagrams. The SMU can perform the
following operations.
•
Applies voltage and measures current
•
Applies current and measures voltage
•
Applies voltage and measures voltage
•
Applies current and measures current
Figure 2-1
Simplified SMU Circuit Diagram
High Force
Chassis ground
V
A
+
-
+
-
High Sense
4-wire connections
Low Sense
Low Force
High Force
Chassis ground
V
A
+
-
+
-
Low Force
2-wire connections