Keysight B2900 User’s Guide, Edition 5
4-31
Front Panel Reference
Config key group
Wait Control dialog box
This dialog box provides the following parameters for setting the source and
measurement wait time.
The source wait time is defined as the time the source channel cannot change the
output after the start of a DC output or the trailing edge of a pulse.
The measurement wait time is defined as the time the measurement channel cannot
start measurement after the start of a DC output or the trailing edge of a pulse.
Ch
Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
State
Wait time, ON or OFF
Automatic
Automatic wait time, ON or OFF
Gain and Offset
Parameters for calculating the wait time. See the following
formula.
•
If State=ON and Automatic=ON:
wait time =
Gain
initial wait time +
Offset
•
If State=ON and Automatic=OFF:
wait time =
Offset
•
If State=OFF:
wait time=0
The initial wait time is automatically set by the instrument, and cannot be changed.