2-10
Keysight B2900 User’s Guide, Edition 5
Introduction
Source/Measure Unit
Measurement Parameters
B2900 supports the following measurement parameters.
•
Current
•
Voltage
•
Resistance
•
Power
Resistance data is given by Resistance=Vmeas/Imeas.
Power data is given by Power=Vmeas
Imeas.
In the above formula, Vmeas is the voltage measurement data, and Imeas is the
current measurement data.
For using the resistance compensation, see
“Resistance Compensation” on page
Limit/Compliance
B2900 has a compliance feature that limits the output voltage or current to prevent
damage to the device under test. When the SMU applies voltage, you can specify the
current limit/compliance. When the SMU applies current, you can specify the
voltage limit/compliance. For details about limit/compliance, see