5-30
Keysight B2900 User’s Guide, Edition 5
Front Panel Operations
To Execute Limit Test
OUT: Failure if the channel comes out of the compliance state.
•
In the Fail Pattern field, set a bit pattern that represents the limit test fail state
(Fail).
9. When the test mode is LIMIT and the operation mode is GRADING, configure
the following settings:
•
In the Up Pattern field, set a bit pattern that represents the
failed-by-exceeding-upper-limit state.
•
In the Low Pattern field, set a bit pattern that represents the
failed-by-exceeding-lower-limit state.
•
In the Up Limit field, set the upper limit for the pass/fail judgement.
•
In the Low Limit field, set the lower limit for the pass/fail judgement.
10. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
NOTE
For using MATH in the pass/fail judgement of the limit test, see
.
For using the resistance compensation, see
“To Enable or Disable Resistance
.