8163A/B, 8164A/B, 8166A/B User’s Guide
185
Return Loss Measurement
7
Figure 103
Return Loss Measurement Setup - Internal Source used
• If an external source is used, connect it to the Return Loss module
Input.
• The measurement patchcord will have an angled connector at the
return loss module. The other end should be straight-polished or
angle-polished to match the input of the devices to be measured.
Remove electrical offsets
1 Make sure that the source is not active and that you have covered the
end of the patchcord to prevent light being coupled into the end. Move
to the Return Loss module channel and press [Zero] to remove
electrical offsets in the instrument.
Set Averaging Time
2 Move to the [
AvgTime
] parameter (the measurement averaging time).
Make sure that the selected averaging time is suitable for your
measurements.
Use an averaging time of at least 200 ms, but increase this to at least 1 s
for return losses greater than 50 dB. Longer averaging times give more
accurate results, but reduce the speed at which the instrument can
complete a measurement.
8161x
8161x
8163B Lightwave Multimeter
NOTE
If you are using an external source, you must fix its output cable to
ensure stability.
Содержание 8163A
Страница 17: ......
Страница 50: ...8163A B 8164A B 8166A B User s Guide 49 Getting Started 1 Figure 19 The 8163B System Configuration Menu...
Страница 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Страница 99: ......
Страница 133: ......
Страница 137: ...136 8163A B 8164A B 8166A B User s Guide 5 Tunable Lasers Figure 77 Setting High Power parameters...
Страница 213: ......
Страница 264: ...8163A B 8164A B 8166A B User s Guide 263 Applications 10 Figure 161 The 8164A B Logging Setup Screen...
Страница 271: ...270 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 166 The 8164A B Logging Analysis Screen...
Страница 277: ...276 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 170 The Keysight 8164A B Stability Setup Screen...
Страница 284: ...8163A B 8164A B 8166A B User s Guide 283 Applications 10 Figure 175 The Keysight 8164A B Stability Analysis Screen...
Страница 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Страница 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Страница 313: ......
Страница 345: ......
Страница 453: ......