8163A/B, 8164A/B, 8166A/B User’s Guide
8
Setting Attenuation and/or Power Levels
How to Set Attenuation
Before applying an attenuation factor, set the Attenuator module to the
appropriate wavelength.
8157xA modules
The instrument utilizes a filter technology with a very low wavelength
dependence. Nevertheless, for the best possible accuracy:
• If the source is of a particular wavelength, set the Attenuator module to
that wavelength.
• If the source is a multi-wavelength DWDM signal, set the Attenuator
module to the weighted wavelength of the spectrum.
on page -220 shows an attenuation factor
α
of 4.000 dB
applied to a 1550nm source.
Figure 135
Attenuation factor set to 4.000 dB
Содержание 8163A
Страница 17: ......
Страница 50: ...8163A B 8164A B 8166A B User s Guide 49 Getting Started 1 Figure 19 The 8163B System Configuration Menu...
Страница 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Страница 99: ......
Страница 133: ......
Страница 137: ...136 8163A B 8164A B 8166A B User s Guide 5 Tunable Lasers Figure 77 Setting High Power parameters...
Страница 213: ......
Страница 264: ...8163A B 8164A B 8166A B User s Guide 263 Applications 10 Figure 161 The 8164A B Logging Setup Screen...
Страница 271: ...270 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 166 The 8164A B Logging Analysis Screen...
Страница 277: ...276 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 170 The Keysight 8164A B Stability Setup Screen...
Страница 284: ...8163A B 8164A B 8166A B User s Guide 283 Applications 10 Figure 175 The Keysight 8164A B Stability Analysis Screen...
Страница 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Страница 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Страница 313: ......
Страница 345: ......
Страница 453: ......