8163A/B, 8164A/B, 8166A/B User’s Guide
3
Power Measurement
To set the power unit to dBm:
1 Move to the [
P
] parameter and press the [Power Unit] softkey. You will
see the menu in
Figure 57
Selecting the Power Unit
2 Move to <
dBm
>, by using the cursor key, and press
Enter
.
How to Set the Calibration Offset
This is a calibration offset that you can enter to compensate for external
optical circuitry, for example, the 81022FF Integrating Sphere or a 3 dB
coupler. The calibration offset, [
Cal
], is automatically subtracted from the
input signal.
P
measured
(dBm) = P
input
(dBm)
-
CAL(dB)
Where,
P
measured
is the adjusted value of the signal read in dBm,
P
input
is the input signal level in dBm, and
CAL
is the calibration offset in decibels (dB).
Содержание 8163A
Страница 17: ......
Страница 50: ...8163A B 8164A B 8166A B User s Guide 49 Getting Started 1 Figure 19 The 8163B System Configuration Menu...
Страница 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Страница 99: ......
Страница 133: ......
Страница 137: ...136 8163A B 8164A B 8166A B User s Guide 5 Tunable Lasers Figure 77 Setting High Power parameters...
Страница 213: ......
Страница 264: ...8163A B 8164A B 8166A B User s Guide 263 Applications 10 Figure 161 The 8164A B Logging Setup Screen...
Страница 271: ...270 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 166 The 8164A B Logging Analysis Screen...
Страница 277: ...276 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 170 The Keysight 8164A B Stability Setup Screen...
Страница 284: ...8163A B 8164A B 8166A B User s Guide 283 Applications 10 Figure 175 The Keysight 8164A B Stability Analysis Screen...
Страница 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Страница 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Страница 313: ......
Страница 345: ......
Страница 453: ......