128
8163A/B, 8164A/B, 8166A/B User’s Guide
4
Laser Sources
How to Set the Frequency of a Modulated Signal
To set the frequency of the modulation to 6.500 kHz:
1 Move to the frequency parameter, [
Mod Frq
], press
Enter
.
2 Enter 6.500 press
Enter
.
How to Set the Modulation Mode
3 Move to the modulation source parameter, [
Mod Src
], and press
Enter
.
4 Move to <
Internal
>, by using the cursor key, and press
Enter
. The text
Int
appears in the Laser Source channel.
How to Increase Linewidth
You can use coherence control to increase the linewidth of the optical
signal output from your 8163A/B Series Laser Source module.
Enabling the coherence control increases the linewidth of the optical
output signal to between 50 and 500 MHz (typically). Coherence control
greatly reduces interference effects and therefore improves the power
stability in sensitive test setups.
To enable coherence control:
1 Move to the Laser Source channel and press [Details].
2 Move to [
Mod Src
] and press
Enter
.
3 Move to <
Coherence Ctrl.
>, Coherence Control, to increase linewidth.
For typical values, refer to User’s Guide for your laser source module,
among the Supplementary Performance Characteristics provided with the
module’s Specifications.
4 Press
Enter.
The text
CC
appears in the Laser Source channel.
How to Modulate Dual-Wavelength Laser Source Modules
NOTE
You cannot modulate both wavelengths of the 81554SM
Dual-Wavelength Laser Source module independently. If you select both
wavelength outputs using the [
λ
] parameter, you can choose to
modulate both outputs at the same frequency or to output both outputs
as CW signals.
Содержание 8163A
Страница 17: ......
Страница 50: ...8163A B 8164A B 8166A B User s Guide 49 Getting Started 1 Figure 19 The 8163B System Configuration Menu...
Страница 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Страница 99: ......
Страница 133: ......
Страница 137: ...136 8163A B 8164A B 8166A B User s Guide 5 Tunable Lasers Figure 77 Setting High Power parameters...
Страница 213: ......
Страница 264: ...8163A B 8164A B 8166A B User s Guide 263 Applications 10 Figure 161 The 8164A B Logging Setup Screen...
Страница 271: ...270 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 166 The 8164A B Logging Analysis Screen...
Страница 277: ...276 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 170 The Keysight 8164A B Stability Setup Screen...
Страница 284: ...8163A B 8164A B 8166A B User s Guide 283 Applications 10 Figure 175 The Keysight 8164A B Stability Analysis Screen...
Страница 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Страница 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Страница 313: ......
Страница 345: ......
Страница 453: ......