200
8163A/B, 8164A/B, 8166A/B User’s Guide
7
Return Loss Measurement
Figure 122
The Calibration Parameters Screens - Return Loss Diode
2 Press [Next]. The
Calibration parameters
screen for the
Monitor Diode
appears, as displayed in
on page -201. This screen shows
current value for the following quantities:
•
[
Para
], the parasitic power value measured by the Return Loss
module’s monitor diode in dBm. This value is determined by the
termination calibraion, or you can use the default setting held in the
factory calibration.
•
[
Ref
], the power measured by the Return Loss module’s monitor
diode during the reflectance calibration in dBm
•
[
Meas
], the power currently measured by the Return loss Module’s
internal monitor diode in dBm
Содержание 8163A
Страница 17: ......
Страница 50: ...8163A B 8164A B 8166A B User s Guide 49 Getting Started 1 Figure 19 The 8163B System Configuration Menu...
Страница 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Страница 99: ......
Страница 133: ......
Страница 137: ...136 8163A B 8164A B 8166A B User s Guide 5 Tunable Lasers Figure 77 Setting High Power parameters...
Страница 213: ......
Страница 264: ...8163A B 8164A B 8166A B User s Guide 263 Applications 10 Figure 161 The 8164A B Logging Setup Screen...
Страница 271: ...270 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 166 The 8164A B Logging Analysis Screen...
Страница 277: ...276 8163A B 8164A B 8166A B User s Guide 10 Applications Figure 170 The Keysight 8164A B Stability Setup Screen...
Страница 284: ...8163A B 8164A B 8166A B User s Guide 283 Applications 10 Figure 175 The Keysight 8164A B Stability Analysis Screen...
Страница 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Страница 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Страница 313: ......
Страница 345: ......
Страница 453: ......