220
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4
Measurement Operations & Settings
Node A Input Configuration
This
32-bit unsigned integer
determines how inputs to Node A are generated.
Bit
31
30
29
18-2
8
17
16
2
1
0
Mnemonic
Logic: 0 for OR, 1 for AND
Input Trigger Connector: 0 - Inactive, 1 - Trigger at Input Trigger Connector can trigger Node A
Node B: 0 - Inactive, 1 - Trigger at Node B can trigger Node A
Not used.
Slot 17: 0 - Inactive, 1 - Event at slot 17 can trigger Node A
Slot 16: 0 - Inactive, 1 - Event at slot 16 can trigger Node A
Slot 2: 0 - Inactive, 1 - Event at slot 2 can trigger Node A
Slot 1: 0 - Inactive, 1 - Event at slot 1 can trigger Node A
Slot 0: 0 - Inactive, 1 - Event at slot 0 can trigger Node A
:TRIGger[n][:CHANnel[m]]:OUTPut
on page 213 explains how slot events can generate triggers.
Hexadecimal
#H
80000000
#H40000000
#H20000000
0
#H20000
#H10000
#H4
#H2
#H1
....
....
Node B Input Configuration
This
32-bit unsigned integer
determines how inputs to Node B are generated.
Bit
31
30
29
18-2
8
17
16
2
1
0
Mnemonic
Logic: 0 for OR, 1 for AND
Input Trigger Connector: 0 - Inactive, 1 - Trigger at Input Trigger Connector can trigger Node B
Node A: 0 - Inactive, 1 - Trigger at Node A can trigger Node B
Not used.
Slot 17: 0 - Inactive, 1 - Event at slot 17 can trigger Node B
Slot 16: 0 - Inactive, 1 - Event at slot 16 can trigger Node B
Slot 2: 0 - Inactive, 1 - Event at slot 2 can trigger Node B
Slot 1: 0 - Inactive, 1 - Event at slot 1 can trigger Node B
Slot 0: 0 - Inactive, 1 - Event at slot 0 can trigger Node B
:TRIGger[n][:CHANnel[m]]:OUTPut
on page 213 explains how slot events can generate triggers.
Hexadecimal
#H
80000000
#H40000000
#H20000000
0
#H20000
#H10000
#H4
#H2
#H1
....
....