8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
213
Measurement Operations & Settings
4
command:
:TRIGger[
n
][:CHANnel[
m
]]:OUTPut
syntax:
:TRIGger[
n
][:CHANnel[
m
]]:OUTPut
description:
Specifies when an output trigger is generated and arms the module.
parameters:
DISabled:
AVGover:
MEASure:
MODulation:
STFinished:
SWFinished:
SWSTarted:
Never.
When averaging time period finishes.
When averaging time period begins.
For every leading edge of a digitally-modulated (TTL) signal
When a sweep step finishes.
When sweep cycle finishes.
When a sweep cycle starts.
response:
none
example:
trig1:outp dis
affects:
All tunable laser modules (except 81950A), series power meters, and return loss modules.
dual sensors:
Can only be sent to master channel, slave channel is also affected.
In continuous mode,
wav:swe:step:[widt]
is used for triggering, see
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:[WIDTh]
command:
:TRIGger[
n
][:CHANnel[
m
]]:OUTPut?
syntax:
:TRIGger[
n
][:CHANnel[
m
]]:OUTPut?
description:
Returns the condition that causes an output trigger.
parameters:
none
response:
DISabled:
AVGover:
MEASure:
MODulation:
STFinished:
SWFinished:
SWSTarted:
Never.
When averaging time period finishes.
When averaging time period begins.
For every leading edge of a digitally-modulated (TTL) signal
When a sweep step finishes.
When sweep cycle finishes.
When a sweep cycle starts.
example:
trig1:outp?
->
dis<END>
affects:
All tunable laser modules (except 81950A), power meters, and return loss modules.
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.