8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4
Measurement Operations & Settings
example:
sens1:func:par:logg?
->
+64,+1.00000000E-001<END>
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
NOTE
Setting parameters for the MinMax function sets some parameters, including
hidden parameters, for the stability and logging functions and vice versa. You
must use the
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax
command to set parameters before you start a MinMax function using the
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe
on page 104 command.
command:
:SENSe[
n
][:CHANnel[
m
]]:FUNCtion:PARameter:MINMax
syntax:
:SENSe[
n
][:CHANnel[
m
]]:FUNCtion:PARameter:MINMax<wsp>
CONTinous|WINDow|REFResh,<data points>
description:
Sets the MinMax mode and the number of data points for the
MinMax data acquisition function.
parameters:
CONTinous:
WINDow:
REFResh:
continuous MinMax mode
window MinMax mode
refresh MinMax mode
Data Points is the number of samples that are recorded in the memory buffer used by the WINDow and REFResh modes. Data
Points is an
integer
value.
See Chapter 3 of the 8163A/B Lightwave Multimeter, 8164A/B Lightwave Measurement System, & 8166A/B Lightwave
Multichannel System User’s Guide, for more information on MinMax mode.
See
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe
on page 104 for information on starting/stopping a data
acquisition function.
See
:SENSe[n][:CHANnel[m]]:FUNCtion:RESult?
on page 100 for information on accessing the results of a
data acquisition function.
See
on page -208 for information on how triggering affects data acquisition functions.
response:
none
example:
sens1:func:par:minm WIND,10
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also affected.