8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
59
Instrument Setup and Status
3
parameters:
The bit value for the OSESM as a
16-bit signed integer
value (0 .. +32767)
The default value is 0.
response:
none
example:
stat:oper:enab:level1 128
command:
:STATus:OPERation:ENABle:LEVel
1
?
syntax:
:STATus:OPERation:ENABle:LEVel
1
?
description:
Returns the OSESM for the OSESR for slots 15 - 17 of the 8166A/B Lightwave Measurement System
parameters:
none
response:
The bit value for the operation enable mask as a
16-bit signed integer
value (0 .. +32767)
example:
stat:oper:enab:level1?
->
+128<END>
command:
:STATus
n
:OPERation[:EVENt]?
syntax:
:STATus
n
:OPERation[:EVENt]?
description:
Returns the Operational Slot Status Event Register (OSSER) of slot
n
.
parameters:
none
response:
The results for the individual slot events (a
16-bit signed integer
value, where 0
£
value
£
32767):
Bit
8-15
7
6
5
4
3
2
1
0
Mnemonic
Not used
Slot
n
: offset (
λ
) type bit 2
Slot
n
: offset (
λ
) type bit 1
Slot
n
: offset (
λ
) has been enabled
Slot
n
: shutter has been opened
Slot
n
: Zeroing ongoing
Not used
Slot
n
: Coherence Control has been switched on
Slot
n
: Laser has been switched on
Decimal Value
0
128
64
32
16
8
0
2
1
example:
stat1:oper?
->
+0<END>
command:
:STATus
n
:OPERation:CONDition?
syntax:
:STATus
n
:OPERation:CONDition?
description:
Returns the Operational Slot Status Condition Register of slot
n
.