8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
181
Measurement Operations & Settings
4
command:
:SOURce<
SlotNr
>:TRANsmitter:TCHeck?
syntax:
:SOURce<
SlotNr
>:TRANsmitter:TCHeck?
description:
Get the temperature check status of the transmitter reference module.
parameters:
none
response:
A value ranging from 0.0 (recalibration necessary) to 1.0.
Values near 1.0 indicate the current temperature is close to the temperature at the last recalibration.
Values near 0.0 indicate the current temperature has drifted towards its maximum tolerance limit to trigger the UNCAL bit.
example:
:SOUR1:TRAN:TCH?
−>
1.0<END>
affects:
81490A