8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
35
Specific Commands
2
:SENSe[
n
][:CHANnel[
m
]]:CORRection
[:LOSS][:INPut][:MAGNitude]/?
Sets or returns the value of correction data for a sensor.
:COLLECT:ZERO
Executes a zero calibration of a sensor module.
:COLLECT:ZERO?
Returns the current zero state of a sensor module.
:COLLECT:ZERO:ALL
Executes a zero calibration of all sensor modules.
:SENSe[
n
][:CHANnel[m]]:FUNCtion
:PARameter:LOGGing/?
Sets or returns the number of samples and the averaging time, t
avg
, for
logging.
:PARameter:MINMax/?
Sets or returns the minmax mode and the window size.
:PARameter:STABility/?
Sets or returns the total time, delay time and the averaging time, t
avg
, for
stability.
:RESult?
Returns the data array of the last function.
:RESult:BLOCk?
Returns a specified binary block from the data array for the last power meter
data acquisition function.
:RESult:MAXBlocksize?
Returns the maximum block size for power meter data acquisition functions.
:RESult:MONitor?
For return loss module, returns monitor diode data array of last function.
:STATe/?
Enables/disables the function mode or returns whether the function mode is
enabled.
:THReshold/?
Sets or returns the threshold value and the start mode.
:SENSe[
n
][:CHANnel[
m
]]:POWer
:ATIMe/?
Sets or returns the average time of a sensor.
:RANGe[:UPPer]/?
Sets or returns the most positive signal entry expected for a sensor.
:RANGe:MONitor[:UPPer]/?
Sets or returns the range of the monitor diode within a return loss module.
:RANGe:AUTO/?
Sets or returns the range of a sensor to produce the most dynamic range
without overloading.
:REFerence/?
Sets or returns the reference level of a sensor.
:UNIT/?
Sets or returns the units used for absolute readings on a sensor.
:WAVelength/?
Sets or returns the wavelength for a sensor.
Command
Description
Page