8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
117
Measurement Operations & Settings
4
command:
:SENSe[
n
]:[CHANnel[
m
]]:POWer:WAVelength?
syntax:
:SENSe[
n
]:[CHANnel[
m
]]:POWer:WAVelength?[<wsp>MIN|MAX|DEF]
description:
Inquires the current sensor wavelength.
parameters:
none
Also allowed are:
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the
minimum programmable value and the maximum programmable value
response:
The wavelength as a
float
value in meters.
example
sens1:pow:wav?
->
+1.55000000E-006<END>
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[
n
]:[CHANnel[
m
]]:RETurnloss:CALibration:FACTory
syntax:
:SENSe[
n
]:[CHANnel[
m
]]:RETurnloss:CALibration:FACTory
description:
For all sources, overwrites the current calibration values with the factory-set calibration settings. See
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
on page 118 and
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]?
on page 121 for information on
calibrating your return loss module.
parameters:
none
response:
none
example
sens1:ret:cal:fact
affects:
All return loss modules
command:
:SENSe[
n
]:[CHANnel[
m
]]:RETurnloss:CALibration:COLLect:REFLectance
syntax:
:SENSe[
n
]:[CHANnel[
m
]]:RETurnloss:CALibration:COLLect:REFLectance
description:
For the currently selected source, start the calibration and save the calibration values for a defined reflectance reference
measurement. See
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]
information on setting the return loss value of your reference reflector.
parameters:
none