94
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4
Measurement Operations & Settings
command:
:SENSe[
n
]:[CHANnel[
m
]]:CORRection[:LOSS][:INPut][:MAGNitude]
syntax:
:SENSe[
n
]:[CHANnel[
m
]]:CORRection[:LOSS][:INPUT][:MAGNitude]<wsp>
<value>[DB|MDB]
description:
Enters a calibration value for a module.
parameters:
The calibration factor as a
float
value
If no unit type is specified, decibels (dB) is implied.
response:
none
example:
sens1:corr 10DB
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[
n
]:[CHANnel[
m
]]:CORRection[:LOSS][:INPut][:MAGNitude]?
syntax:
:SENSe[
n
]:[CHANnel[
m
]]:CORRection[:LOSS][:INPUT][:MAGNitude]?
description:
Returns the calibration factor for a module.
parameters:
none
response:
The calibration factor as a
float
value. Units are in dB, although no units are returned in the response message.
example:
sens1:corr?
->
+1.00000<END>
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[
n
]:[CHANnel[
m
]]:CORRection:COLLect:ZERO
syntax:
:SENSe[
n
]:[CHANnel[
m
]]:CORRection:COLLect:ZERO
description:
Zeros the electrical offsets for a power meter or return loss module.
parameters:
none
response:
none