8.11 Automatic taring
This function is to be used for faster calculation of the net weight, in case the tare
load changes for each weighing.
Operation:
Display:
Ö
Turn on display unit by pressing the
ON/OFF
key
0.00 g
Ö
Press
F
-key
1 Funi
Ö
Repeatedly press the
a
key until
6 AtAr
appears
6 AtAr
Ö
Press
TARE
key
0.00 g
Ö
When the weighing plate is without load, press the
a
key
0.00 g
Ö
Put vessel of goods to be weighed on weighing plate
123.45 g
Ö
A stable weighing value will start automatic taring
------
Ö
The display shows
NET 0.00 g
Ö
Place weight in weighing container, net weight of weighed
item will be shown.
NET 67.23 g
Ö
Remove weight together with weighing container
0.00 g
Ö
Set up weighing container etc. for next weighing process
Ö
Go back to weighing mode by pressing the
F
key twice.
0.00 g
Important
: Ensure that the minimum weight
entered (For setting see chpt. 9.5) is lower than the weight of the weighing container
or the weighing container will not be tared automatically.