KLB-TM-BA-e-0811
53
13.6 Calibrated display unit P 0.b LFt
This parameter was included because of the verifiable display units, as these have
some restrictions. This way, access to illegal functions is barred The edition of
unstable weighing values and user adjustment become impossible.
Settings
O.b LFt
:
YES : verifiable
No.: non-verifiable
The identification of verifiable and non-verifiable display units is indicated
immediately after start-up by showing the version number.
Program version for non-verifiable display unit:
tcnP x.x
Program version for verifiable display unit:
tcLP x.x
13.7 Selecting the power supply P 0.d Acu
This parameter is used to set the available power supply.
No
.:
No Bat-Lo display
SLA
:
6V battery block, Bat-Lo display appears when voltage drops below 5.7 V.
nimh
: 6 x 1.2V battery block (type AA) , Bat-Lo display appears when voltage
drops below 7.2 V.
Attention
: When 6 x 1.5 V batteries are used, it is essential that the parameter 5.5
CHr6 is set to “NO”. This switches off the charging process as otherwise the
batteries and possibly even the display unit might be destroyed.
Notice
: During configuration it is important that this parameter is set according to
the selection above. A drop in voltage is indicated by a voltage monitor
displaying the battery symbol and the subsequent switch-off of the balance
according to the used storage battery type. All voltage limits for the various
storage batteries are determined by setting this parameter.
This parameter is ineffective if the display unit is powered by the plug-in power unit.