Notice:
Display unit is equipped with rechargeable NiMH batteries and plug-in power pack.
During initial operation it is very important to charge the batteries for c. 12 hours.
Afterwards discharge the batteries completely 3 times (watch display message and
automatic disconnection) and recharge. If this is done, the life of the batteries will be
increased and the nominal capacity of the batteries reached.
8.7 P1 rEAd – basic settings
8.7.1 Filter settings
This menu item allows the display unit to be set according to specific ambient
conditions and measuring purposes.
OFF Filter turned off
1-4 Filter
levels:
1
4
Sensitive and fast (very quiet set-up location).
Robust but slow (very busy set-up location)
Operation:
Display:
Ö
Selecting the menu item
P1 rEAd
Ö
Press
TARE
key
1.1.Fil
Ö
Press
TARE
key
1
Ö
To change your selection, press the
a
key.
OFF
Ö
1
Ö
2
Ö
3
Ö
4
Ö
Setting will be imported by pressing the
key.
1.1.Fil
KLB-TM-BA-e-0811
27