13.3 Calling menu for entering parameter of configuration
Turn on display unit by pressing the
ON/OFF
key and simultaneously press the micro
switch
S1
(for approximately 5 seconds) (see image below).
After the auto test of the balance, press the
F
and the
key simultaneously.
After the menu was called, select
P.0 FAct
by pressing the
a
key.
For navigation around menu refer to chapter 8.2.
To call specific parameters, press the
TARE
key. To change the parameter, press
the
a
key.
To store, press the PRINT key; to exit without storing, press the
F
key.
Notice
: After making and storing changes, turn the display unit off, then turn it on
again.
Micro switch S1
After carrying out the configuration, close the case and insert the screws.
KLB-TM-BA-e-0811