![background image](http://html1.mh-extra.com/html/kern/klb-tm/klb-tm_operating-manual_1957000049.webp)
13.2 Parameter description for configuration
Parameter
Ground
setting
Range
Description
P 0.0 A/d
-
-
A/D converter value
P 0.1 Uni
g
kg - lb - g
Weighing unit
P 0.2 div1
0.001
0.000 - 50
Lowest value digit and zero position of
first range
P 0.3 div2
0.001
0.000 - 50
Lowest value digit and zero position of
second range
P 0.4 FulS
006.009
-
Maximum load
P 0.5 rn 2
000.000
-
Switch point between first and second
range
P 0.6 Auto
0.25 d
0.10 d–
5.00 d
Auto zero range
P 0.7 wEi
000.500
0 - 1
Adjustment weight in respect of
maximum load
P 0.8 St_u
-
-
Initial weight for adjustment
P 0.9 CAL
-
-
Default adjustment
P 0.A Gcor 1.00000
0.90000 –
1.10000
Factor for gravity constant
P 0.b LFt
no
no - yes
Verified display unit
P 0.c rAn
yes
no - yes
Initial weight adjustment
yes – ON, no – OFF
P 0.d Acu
no
no – SLA-
NiMH
Selecting power supply
No- without Bat-Lo display
SLA- battery pack 6V with Bat-Lo display
NiMH-batteries 6x 1.2V with Bat-Lo
display
P 0.E toSC
IndSt
IndSt-
HunnA-
Hunnb
Firmware version:
IndSt Industrial balance
HunnA Medical balance (Hold + BMI)
Hunnb Medical balance (Hold)
P 0.F dFLt
-
-
Reset to default setting
P 0.L Line
-
-
Linearization (max. 6 points)
KLB-TM-BA-e-0811
49