KLP 073008
4-7
3. The status display reads
INPC
. Connect the reference voltage used to define full scale out-
put current as described in PAR 3.7
.
4. When the reference voltage is at the desired value, click the
PRESS TO ACCEPT
button
to
accept the value. When the calibration is saved, this voltage value will be equivalent to full
scale output current.
NOTE: The full scale output current is determined by the virtual
settings in effect.
5. The status display will not change. Proceed to PAR. 4.4.4 to exit calibration.
4.4.4
CALIBRATION EXIT (VISA DEMO)
NOTE: Clicking the Calibrate
CLOSE
button on the main panel will exit calibration WITHOUT
saving any calibration data.
1. Click the
SAVE CALIBRATION
button on the calibration window to save this calibration.
2. The Calibration window automatically closes and the Virtual Model window automatically
opens.
3. The virtual model settings can be changed after the password has been properly entered.
NOTE: The virtual model is always reset following calibration. If voltage calibration is com-
pleted (regardless of whether or not current calibration is completed), the virtual model
voltage is set for maximum rated voltage, and current is set to the value corresponding
to 1200 Watts. If a current calibration is completed, but no Voltage calibration, the vir-
tual model current will be set for maximum rated current and voltage is set to the value
corresponding to 1200 Watts.
4.5
CHANGING THE CALIBRATION PASSWORD
The 4-digit password is required for to enter calibration mode. The factory default password is
listed in Table 4-1. To change the password from the front panel proceed as follows:
1. If the status display shows
SET
, tap either the
CURRENT
or
VOLTAGE
controls to take the
unit out of setpoint mode (status display goes from
SET
to blank)
2. Hold the
PROTECT
switch in and use a thin tool (e.g., a paper clip), to press the
FUNCTION
switch. Status display reads
PASS.
NOTE: To exit this procedure without changing the password, rotate the
VOLTAGE
control.
3. Tap the
DC OUTPUT
switch. Status display reads
OLD?
.
4. Rotate the
CURRENT
control until
DC AMPERES
display shows old password. Tap the
DC
OUTPUT
switch to enter. Status display shows
NEW?.
5. Rotate the
CURRENT
control until
DC AMPERES
display shows new password. Tap the
DC
OUTPUT
switch to enter. Status display shows
RPT?
.
6. Rotate the
CURRENT
control until
DC AMPERES
display reads new password. Tap the
DC
OUTPUT
switch to enter. Status display reads
DONE
and the unit returns to setpoint mode.
Содержание KLP SERIES
Страница 2: ......
Страница 6: ......
Страница 14: ...viii KLP073008 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 20: ...1 6 KLP 073008 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...
Страница 21: ...KLP 073008 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ......
Страница 40: ......
Страница 86: ......