4-2
KLP 073008
will be connected across the output terminals and the digital voltmeter will be connected across
the sense terminals of the shunt resistor.
1. Turn off power supply and disconnect load from output terminals at the rear of the unit.
2. Verify that power supply is configured for local error sensing (PAR. 2.7.5.1).
3. For voltage calibration connect DVM to (M+,S+) and (M-,S-) of terminal block at the rear of
power supply.
4. For current calibration connect an appropriate shunt resistor (see PAR 4.2) across the output
terminals and connect DVM to shunt resistor terminals.
5. Turn on power supply, exit
SET
mode by tapping either
VOLTAGE
or
CURRENT
control and
using a thin tool (e.g., a paper clip), press
FUNCTION
switch repeatedly until status displays
reads
UTIL
. Rotate either
VOLTAGE
or
CURRENT
control until status display reads
CAL
.
6. Tap
DC OUTPUT
switch to enter calibration mode. Status display reads "
PASS
" and
CUR-
RENT DISPLAY
reads
0000
while the unit waits for the calibration password to be entered.
The leftmost digit will be flashing.
7. Rotate
CURRENT
control to select a number from 0-9. Tap
CURRENT
control to advance to
next digit. Repeat this process until all four digits have been selected. Tap
DC OUTPUT
switch to accept the number.
8. If password is incorrect, the status display reads "
ERR
". Repeat step 7 to enter the correct
password, or press the
FUNCTION
switch to exit calibration
9. If password is correct, status display reads
V ?
. Tapping either
VOLTAGE
or
CURRENT
control continuously cycles through all the calibration choices. The status display reads:
V ?
for voltage calibration
C ?
for current calibration,
EXT?
for remote analog calibration, and
EXIT
to save the calibrated results. Tap the
FUNCTION
switch to exit calibration WITHOUT
saving any calibration data (see PAR. 4.3.4). Tap the
DC OUTPUT
switch to select the dis-
played choice.
NOTE: Calibrations may be done in any order.
10.Proceed to PAR. 4.3.1 for voltage calibration, PAR 4.3.2 for current calibration, PAR. 4.3.3
for external calibration or PAR. 4.3.4 to exit calibration
4.3.1
VOLTAGE CALIBRATION (LOCAL)
During voltage calibration, the voltage, voltage readback and remote analog voltage readback
are calibrated.
1. Minimum voltage is first; the status display reads
V_0
. Monitor DVM and rotate the
VOLT-
AGE
control to increase or decrease the output voltage until the DVM reads as close as pos-
sible to minimum rated voltage.
2. Tap the
DC OUTPUT
switch to accept the value. The status display then reads
VMAX
. Moni-
tor DVM and rotate the
VOLTAGE
control to increase or decrease the output voltage until
the DVM reads as close as possible to maximum rated voltage.
Содержание KLP SERIES
Страница 2: ......
Страница 6: ......
Страница 14: ...viii KLP073008 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 20: ...1 6 KLP 073008 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...
Страница 21: ...KLP 073008 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ......
Страница 40: ......
Страница 86: ......