4-4
ATE-DMG 020613
1. If the unit is not in command entry status , press
CLEAR
or
RESET
to put the unit in com-
mand entry status.
2. Press
MENU
key five times. LCD reads
Change Passw. OldPass
.
3. Enter the old password and press
ENTER
. Then enter the new 4-digit Password and press
ENTER
.
4. The new password is accepted and the LCD shows the next Menu screen
GoTo Previous
CALIB? 1 Yes
.
5. Press
CLEAR
to exit the Menu, press
1
to restore the previous calibration values (see PAR.
4.5), or
MENU
for additional menu screens.
4.5
RESTORING PREVIOUS CALIBRATION VALUES
Each time the unit is calibrated, the previous calibration values are retained in non-volatile mem-
ory. If it is determined that a new calibration is erroneous, the previous calibration values can be
restored.
1. If the unit is not in command entry status , press
CLEAR
or
RESET
to put the unit in com-
mand entry status.
2. Press
MENU
key six times. LCD reads
GoTo Previous CALIB? 1
.
3. Press
1
, then
ENTER
. LCD reads
Calibration ! PassWord
.
4. Enter 4-digit password and press
ENTER
. Previous calibration values are restored.
4.6
RESTORING FACTORY CALIBRATION VALUES
The original factory calibration values can be restored.
1. If the unit is not in command entry status , press
CLEAR
or
RESET
to put the unit in com-
mand entry status.
2. Press
MENU
key seven times. LCD reads
GoTo Factory CALIB? 1
.
3. Press 1, then
ENTER
. LCD reads
Calibration ! PassWord
.
4. Enter 4-digit password and press
ENTER
. Factory calibration values are restored.
Содержание ATE 100-10DMG
Страница 2: ......
Страница 6: ......
Страница 14: ......
Страница 15: ...ATE DMG020613 ix FIGURE 1 1 ATE DMG SERIES PROGRAMMABLE POWER SUPPLY...
Страница 16: ...ix Blank x ATE DMG020613...
Страница 24: ...1 8 ATE DMG OPER 020613 FIGURE 1 2 MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...2 2 ATE DMG 020613 FIGURE 2 1 ATE DMG FULL RACK SERIES FRONT PANEL...
Страница 34: ......
Страница 74: ......
Страница 98: ......