3-14
ATE-DMG 020613
3.4.13
RECALLING STORED OUTPUT SETTINGS
With the power supply in command entry status , press
RECALL
. The LCD reads
RECALL
mem
nn
where nn is the memory location holding the settings to be retrieved. Press
ENTER
or
CLEAR
to exit without changing setting. Enter memory location (from 1 to 40) and press
ENTER
. The stored settings for voltage, current, overvoltage protection and overcurrent protec-
tion replace the current values, and the unit returns to command entry status.
3.4.14
ENABLING REMOTE ANALOG CONTROL
The power supply may be controlled by external analog voltages (see PAR. 3.5), but the feature
must be enabled first. With the power supply in command entry status , press
MENU
key until
LCD reads
Extern. Ref On
or
Extern. Ref Off
and the bottom line reads
0=Off,
1=On
. Enter 1 (enable) or 0 (disable), then press ENTER to enable or disable the feature. Refer
to PAR. 3.5 for using this feature.
3.4.15
LOCAL MODE PROGRAMMING OF THE POWER SUPPLY.
Local mode programming offers the user 40 memory locations that can be used to program the
power supply. Each location defines values for output voltage, output current, overcurrent pro-
tection, overvoltage protection, time duration (between 0.1 and 2.5 seconds) for the pro-
grammed settings, and the address of the next memory location in the program.
NOTE: When programming sequential voltage levels, it is important to set the Overvoltage to
accommodate the highest voltage of the sequence. Otherwise, when going from higher
to lower voltage levels, the overvoltage protection will trip and shut down the unit.
3.4.15.1 CREATING OR MODIFYING A PROGRAM (PROGRAM EDIT MODE)
Creating a program and modifying an existing program are identical except that a program mod-
ification requires only selected parameters to be changed rather than all new parameter values.
With the power supply in command entry status , press
EDIT PROG
key. The top line indicates
the
STARTaddr. nn
, where nn is one of 40 memory location. To select a different address,
enter new value and press
ENTER
.
The top line of the LCD indicates the active memory location, e.g.
ViewVAL Mem <n>
where n
is the memory location previously chosen. The bottom line of the LCD indicates the parameter
on the left
(Iset
,
Vset
,
OVset
,
OCset
,
TIMEval
,
NEXT STEP
), the parameter’s value on the
right, and data entry status in the middle. To modify time values see PAR.3.4.15.1.1.
Press
ENTER
to accept displayed value, or enter new value and press
ENTER
to accept new
setting (the LCD displays the next parameter). Press
CLEAR
to exit Program Edit mode without
changing value. Use
"
or
ç
key to scroll forward (or backward) to view next (or previous)
parameter or memory location. For a program to run once, then stop, see PAR. 3.4.15.6; to
cycle a program continuously, see 3.4.15.4. After all program values have been set, press
CLEAR
or
ENTER
to exit Program Edit mode. Setting up a program can be made easier by
copying the Memory Location Worksheet (Table 3-3) and filling in the values before program-
ming the power supply.
3.4.15.1.1 MODIFYING PROGRAMMED TIME INTERVAL
The
TIME
key offers a quick and easy way to change the time for any memory location. With the
power supply in command entry status , press
TIME
key. The LCD displays
TIMEval
nn
where nn is the time value set for the selected location. Press
ENTER
to accept displayed
Содержание ATE 100-10DMG
Страница 2: ......
Страница 6: ......
Страница 14: ......
Страница 15: ...ATE DMG020613 ix FIGURE 1 1 ATE DMG SERIES PROGRAMMABLE POWER SUPPLY...
Страница 16: ...ix Blank x ATE DMG020613...
Страница 24: ...1 8 ATE DMG OPER 020613 FIGURE 1 2 MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...2 2 ATE DMG 020613 FIGURE 2 1 ATE DMG FULL RACK SERIES FRONT PANEL...
Страница 34: ......
Страница 74: ......
Страница 98: ......