ATE-DMG 020613
4-3
1. LCD reads
CONNECT DVM TO SHUNT
. Connect Precision Shunt across output (+) and (–)
terminals of power supply and connect DVM to sense terminals of shunt. Press
ENTER
when connections are complete.
NOTE: During calibration the
"
(increase) or
ç
(decrease) keys are used for fine adjust-
ments. If necessary, use the
1
(decrease) and
3
(increase) keys for coarse adjust-
ment, then use the
"
or
ç
keys.
2. With LCD reading
Iout ZERO adj.
, monitor DVM and use
"
or
ç
keys to increase or
decrease output current as necessary to get DVM reading as close to zero as possible (with-
out going to negative value). If unable to reach zero, adjust I
O
ZERO control as necessary.
Press
ENTER
when complete.
3. LCD will flash
Iout F.S.adj. **WAIT**
then
Iout F.S. adj < >
. Monitor DVM and
use
"
or
ç
keys to increase or decrease output voltage as necessary to get DVM reading
as close as possible to value listed below (without exceeding value listed).
NOTE: If LCD reads
CAN’T ADJUST FULL SCALE
, adjust I
O
max control R18 for current
equal to I
O
max. Depress
ENTER
and
RESET
, then restart current calibration.
–
ATE 6-100DMG: 1.000V if 0.01 Ohm resistor, 0.500V if 0.05 Ohm resistor
–
ATE 15-70DMG: 0.700V
–
ATE 25-40DMG: 1.600V
–
ATE 36-30DMG: 1.500V
–
ATE 55-20DMG: 1.000V
–
ATE 75-15DMG: 1.500V
–
ATE 100-10DMG: 1.000V
–
ATE 150-7DMG: 1.400V
4. Press ENTER when complete.
4.4
CHANGING THE CALIBRATION PASSWORD
The 4-digit password is required for to enter calibration mode. The factory default passwords
are listed in Table 4-2.
TABLE 4-2. FACTORY DEFAULT CALIBRATION PASSWORDS
MODEL
PASSWORD
ATE 6-100DMG
6100
ATE 15-50DMG
1550
ATE 25-40DMG
2540
ATE 36-30DMG
3630
ATE 55-20DMG
5520
ATE 75-15DMG
7515
ATE 100-10DMG
1001
ATE 150-7DMG
1507
Содержание ATE 100-10DMG
Страница 2: ......
Страница 6: ......
Страница 14: ......
Страница 15: ...ATE DMG020613 ix FIGURE 1 1 ATE DMG SERIES PROGRAMMABLE POWER SUPPLY...
Страница 16: ...ix Blank x ATE DMG020613...
Страница 24: ...1 8 ATE DMG OPER 020613 FIGURE 1 2 MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...2 2 ATE DMG 020613 FIGURE 2 1 ATE DMG FULL RACK SERIES FRONT PANEL...
Страница 34: ......
Страница 74: ......
Страница 98: ......